          EPA1664—己烷提取物的檢測

          發布時間:2013/6/23      點擊次數:8512

           1) Assemble

          a) Assemble cartridge adapters on the vacuum manifold

          b) Place ECUNIOGXF cartridge(s) on each vacuum station needed

          c) Connect the manifold to a suitable trap and attach the trap to a vacuum system capable of

          attaining a minimum of 25” Hg (635 mm) of vacuum

          d) Insert waste collection vial in manifold (optional)

          2) Prepare Water Sample

          a) Adjust the pH of the sample to less than 2 by adding 5 mL of 6N HCl or 2.5 mL of concentrated

          H2SO4

          b) If acid was added to the sample in the field do not add more unless the pH is greater than 2

          Gloves are recommended as skin oils may affect final sample weight

          3) Condition the Cartridge

          a) Rinse sides of cartridge and bottle holder with 10 mL of hexane

          b) Allow cartridge to soak for 1 minute

          c) Draw the hexane through the cartridge using vacuum

          d) Discard the hexane from waste vial

          e) Draw full vacuum through the cartridge for 2 minutes to dry

          f) Add 10 mL of methanol to the cartridge

          g) Slowly draw the methanol through leaving a layer on the cartridge frit

          h) Soak for one minute

          i) Remove waste vial and discard methanol

          From this point on do not let the cartridge dry out otherwise repeat at step f)

          j) Add 30 mL of DI water to the cartridge

          k) Draw the water through the cartridge to waste

          4) Sample Addition

          a) Add the 1 liter water sample directly to the cartridge

          b) Draw the sample through the cartridge under low vacuum. This may take several minutes

          depending on the solids in the sample. (Note 1) Increase vacuum pressure if necessary. Do

          not exceed 500 mL/minute

          c) Remove the cartridge and tap any excess water from the bottom of the cartridge

          d) Replace cartridge and allow to dry under full vacuum for 10 minutes

          e) Remove any water remaining in the bottom support of the cartridge with a paper towel if

          necessary

          5) Elution

          a) Prepare an extract collection vial containing about 8 mm (0.3 inch) of anhydrous sodium

          sulfate

          b) Place the vial in the manifold station under the cartridge

          c) Rinse the water sample bottle with 10 mL of hexane

          d) Add the hexane to the cartridge

          e) Soak cartridge for 2 minutes

          f) Turn on vacuum and slowly draw the hexane through the cartridge and into the collection vial

          g) Turn off vacuum then repeat steps 5 c) – f) 2 additional times but use 10 mL of hexane

          h) Do not allow the solvent to splash into the collection vial

          i) Add another 10 mL of hexane to the cartridge, rinsing the bottle holder

          j) Soak cartridge for 2 minutes

          k) Draw the hexane through the cartridge

          6) Dry the Extract

          a) Remove the collection vial from the manifold and cover with a screw cap

          b) Shake the extract to form a water/hexane emulsion and immediay pour the extract through a

          sodium sulfate funnel or column containing approximay 40 g of anhydrous sodium sulfate

          held in place with a glass wool plug or frit. Do not use filter paper

          c) Collect the extract in a clean, pre-weighed vessel

          d) Rinse the collection vial with hexane and add it to the sodium sulfate. This will rinse the vial

          and the sodium sulfatez

          7) Gravimetric analysis

          a) Carefully evaporate the hexane using a TurboVap® or similar evaporator at 40° C until the

          extract just reaches dryness.

          Note: Do not over dry or low recoveries will result

          Do not dry on a hot plate or in an oven or over dry as low recoveries may result

          b) Allow to cool to room temperature in a desiccator before weighing

          c) Record this weight as the mass per unit volume of oil and grease and report as HEM as mg/L

          Notes

          1) If very high solids are present, add glass wool to the cartridge prior to extraction to prevent clogging and improve

          flow. The glass wool must be thoroughly rinsed as part of the cartridge during the elution step.

          2) Stearic acid must be in solution in the spiking solution or low recoveries will result. If small crystals are present in

          the spiking solution, sonicate or shake until dissolved.

          DCN-118170-


          文件下載    
          廣州艾欣科學儀器有限公司
          地址:廣東省廣州市天河區元崗街道元崗路天河慧通產業廣場A1棟六樓1604(總部)
          郵箱:fighter31668@163.com
          傳真:86-020-36536169
          關注我們
          歡迎您關注我們的微信公眾號了解更多信息:
          歡迎您關注我們的微信公眾號
          了解更多信息
          主站蜘蛛池模板: 亚洲精品国产美女久久久| 精品亚洲成AV人在线观看| 亚洲乱码中文字幕综合| 农村寡妇一级毛片免费看视频| 99在线视频免费观看| 亚洲国产一区明星换脸| 色噜噜亚洲男人的天堂| 精品国产无限资源免费观看| 久久久久无码专区亚洲av| 亚洲色成人网站WWW永久四虎| 无码国产精品久久一区免费| 国产精品亚洲片在线va| 日本免费网站观看| 亚洲国产成AV人天堂无码| 在线精品一卡乱码免费| 亚洲第一男人天堂| 国产jizzjizz视频免费看| 一级做a爱过程免费视频高清| 在线亚洲午夜理论AV大片| 午夜亚洲乱码伦小说区69堂| 亚洲精品成人区在线观看| 国产啪精品视频网站免费尤物| 亚洲伊人久久综合中文成人网| 亚洲一区二区三区成人网站| 日本免费一区尤物| 中文字幕在线免费播放| 亚洲老熟女@TubeumTV| 久久免费看少妇高潮V片特黄| 亚洲精品WWW久久久久久| 日本免费久久久久久久网站| 亚洲成a人片在线观看中文!!! | 久久精品成人免费网站| 亚洲日本乱码一区二区在线二产线| 免费无码又爽又刺激聊天APP| 无码天堂亚洲国产AV| 亚洲国产精品无码成人片久久| 免费观看激色视频网站bd| 特级aa**毛片免费观看| 国产又长又粗又爽免费视频| 久久久受www免费人成| 亚洲综合伊人制服丝袜美腿|